PART
|
96T
|
48T
|
STORAGE
|
|
Pre-Coated Plate
|
8×12
|
8×6
|
4℃ / -20℃
|
|
Standards
|
2 vials
|
1 vial
|
4℃ / -20℃
|
|
Biotinylated Antibody (1:100)
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Enzyme Conjugate(1:100)
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Enzyme Diluent
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Antibody Diluent
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Standard Diluent
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Sample Diluent
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Washing Buffer (1:25)
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Color Reagent A
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Color Reagent B
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Color Reagent C
|
1 vial
|
1 vial
|
4℃ / -20℃
|
|
Manual
|
1 set
|
1 set
|
RT
|
Place collected whole blood in refrigerator at 4℃ overnight.Then centrifuge for 10mins at 1000-3000 rpm.Take supernatant and either test immediately or place samples at -20℃/-80℃(1-3 months) for storage.
Take plasma where EDTA,sodium citrate,or heparin has been added as anticoagulant. Mix well.Centrifuge mixture for 10mins at 1000-3000 rpm.Take supernatant and test immediately or place samples at -20℃ /-80℃ (1-3 months) for storage.
Take tissue slices which have been washed in 0.01M PBS and then a tissue protein extraction reagent has been added according to proportion of 1g to 5-10mL and then mixed on ice bath. After sufficient homogenization,please then centrifuge for 10mins at 5000-10000rpm.Take supernatant for immediate testing,or place samples at -20℃/-80℃ (1-3 months) for storage.
Centrifuge for 10mins at 1000-3000 rpm.Take supernatant for immediate testing,or place samples at -20℃/-80℃ (1-3 months) for storage.
Centrifuge for 10mins at 1000-3000 rpm.Take supernatant for immediate testing,or place samples at -20℃/-80℃ (1-3 months) for storage.
|
TYPICAL DATA
|
Reference STD1
|
Reference STD2
|
Reference STD3
|
Reference STD4
|
Reference STD5
|
Reference STD6
|
Reference STD7
|
Reference STD8
|
|
Standard Concentration,ng/mL
|
50
|
25
|
12.5
|
6.25
|
3.125
|
1.5625
|
0.78125
|
0
|
|
OD1
|
2.545
|
1.535
|
0.873
|
0.476
|
0.299
|
0.165
|
0.097
|
0.055
|
|
OD2
|
2.596
|
1.489
|
0.881
|
0.485
|
0.296
|
0.165
|
0.097
|
0.053
|
|
AVERAGE
|
2.571
|
1.512
|
0.877
|
0.481
|
0.297
|
0.165
|
0.097
|
0.054
|
STANDARD CURVE
To assess linearity of the assay, samples were spiked with high concentrations of Human Poly Serium Anti-gen(PHSA) and diluted with the appropriate calibrator diluent to produce samples with values within the dynamic range of the assay.
|
Matrix
|
Scope %
|
Average % Recovery Rate
|
|
serum(n=5)
|
86-100
|
93
|
|
EDTA plasma(n=5)
|
91-98
|
94.5
|
|
Cell Culture Media(n=5)
|
91-110
|
100.5
|
|
Within-Run Precision
|
Inter-Batch Precision
|
|||||
|
Sample
|
1
|
2
|
3
|
1
|
2
|
3
|
|
n
|
20
|
20
|
20
|
20
|
20
|
20
|
|
CV%
|
5.4
|
5.9
|
6.8
|
7.8
|
7.9
|
8.6
|